Boonton 4500B RF Peak Power Analyzer
Remote Operation
5-16
CALibration:USER:FREQcal
Description: Performs a single-point sensor gain calibration of the selected sensor using an external (user-
supplied) 0dBm power reference at the current operating frequency. This method can be
combined with an AUTOCAL to yield the most accurate power readings at frequencies that are far

factory-stored values with a measured result. If the channel Frequency setting is changed, the
original cal-factor will be restored. A FREQcal that attempts to set a cal-factor larger than 3 dB
will be rejected and an error returned.
Syntax: CALibration[1|2]:USER:FREQcal ?]
Returns: 0 if successful, 1 otherwise (query form only)
Valid Modes: Power sensors only.
CALibration:ZERO
Description: For a sensor that has been previously auto calibrated and has valid auto-cal status, performs a zero
offset null adjustment.
Syntax: CALibration[1|2]: ZERO[?]
Returns: 0 if successful, 1 otherwise (query form only)
Valid Modes: Power sensors only.
CALibration:ZERO
Description: Performs an offset null adjustment for the specified external trigger input. No external signal can
be applied during calibration.
Syntax: CALibration[3|4]: ZERO[?]
Returns: 0 if successful, 1 otherwise (query form only)
Valid Modes: External trigger inputs only..