Boonton 4500B RF Peak Power Analyzer
Operation
4-100
Filter setting. This results in a "peak hold" characteristic where the min and max peaks increase immediately on the trace
when acquired and then, decay slowly. To hold the peaks longer, increase the Chan # > Exten > Filter setting. To maintain
the min and max peaks, turn Chan # > Exten > Peak Hold hannel.
In Statistical mode (Option 10 only), the CDF and CCDF are drawn filled in below the trace line when enveloping is on. The
PDF is always drawn as a filled distribution.
Note that the envelope setting does not affect any readings returned by markers, automeasure system, or remote commands.
Only the display presentation is affected.
Table 4-44. Display > Menu
Menu Item Function
(Type) Selections (SCPI cmd)
Persistence Pulse Enables or disables trace persistence.
(Toggle) Off, On See discussion above, Sect. 4-18.
DISPlay:PERSistence {ON, OFF}
Envelope Pulse, Mod & Stat Enables or disables trace envelope mode.
(Toggle) Off, On See discussion above, Sect 4-18.
DISPlay:ENVElope {ON, OFF}
Beep Off, On Controls the processor audible beeper
(Toggle) 
is pressed.
SYSTem:BEEP {ON, OFF}
Brightness 10% to 100% Controls the display backlight
(Numeric)
Use any of the data entry keys to vary the
intensity of the LCD backlight. The default
setting is 90%.
DISPlay:BACKlight:BRIGhtness
<0 to 100>
Screen Saver MENU Opens the screen saver submenu
(Submenu)
Fig 4-60. Display >
Menu