Boonton 4500B Peak Power Meter Uživatelský manuál Strana 121

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 372
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 120
Boonton 4500B RF Peak Power Analyzer
Operation
4-49
Table 4-9B. Trig > Menu (modulated)
Menu Item
(Type) Selections Function
Trig Mode Mod Select the trigger system mode of operation
(Toggle) Triggered, Free Run
Press the Trig > Trig Mode menu key to
select the desired mode of trigger system
operation.
Triggered A trigger can be used to Start
the measurement, equivalent to pressing the
Run menu key. The measurement will
Continue until stopped with the menu key
or a bus command.
Free Run Measurements are made
continuously without the need for a
Trigger.
TRIGger:MODE {TRIGGERED, FREERUN}
Trig Source Mod Same as Trig > Trig Source, Table 4-9 above
(Mult. Choice) CH1, CH2,
Ext Trig1, Ext Trig2
Trig Level Mod Same as Trig > Trig Level , Table 4-9 above
(Numeric) See Table 4-9
Trig Slope Mod Same as Trig > Trig Slope, Table 4-9 above
(Toggle) +, -
Fig. 4-27e. Trig > menu
Modulated Mode
Zobrazit stránku 120
1 2 ... 116 117 118 119 120 121 122 123 124 125 126 ... 371 372

Komentáře k této Příručce

Žádné komentáře