Boonton 4500B RF Peak Power Analyzer
Operation
4-33
Table 4-7B. Ext Trig # > Menu (continued)
Menu Item Function
(Type) Selections (SCPI cmd)
Vert. Center Pulse, Mod & Stat Sets the voltage level at the vertical center of the display
(Numeric) Continuous Range:
±1 kV Use this control to shift the waveform display vertically to the desired
position in the window. Press the Ext Trig # > Vert Center menu key
to activate this window. Then use the keypad to shift the vertical center of the display to the
desired level. Increasing the level moves the waveform down; decreasing the level moves it up.
DISPlay:TRACe[3|4]:VCENTer <arg> (volts/Div)
Calibration Pulse, Mod & Stat Accesses the Ext Trig # > Calib > submenu. See Table 4-7C.
(Submenu) MENU
Extensions Pulse, Mod & Sta t Accesses the Ext Trig > Exten > submenu. See Table 4-7D.
(Submenu) MENU