Boonton 4540 Series RF Power Meter
Remote Operation
5-33
MARKer:POSItion:TIMe
Description: Set or return the time (x-axis-position) of the selected marker relative to the trigger. Note that
time markers must be positioned within the time limits of the trace window in the graph display. If
a time outside of the display limits is entered, the marker will be placed at the first or last time
position as appropriate.
Syntax: MARKer[1|2]:POSItion:TIMe <numeric_value>
Argument: <numeric_value> = display-time in seconds (see restrictions)
Valid Modes: Pulse and Modulated Modes.
Restrictions: For zero delay trigger position in the center of the display , the following relationship must be
satisfied: TrigDly - (5*time/div) < MarkerTime < TrigDly + (5*time/div) where the timebase