Boonton 4540 Series RF Power Meter Contents x MEMory:SNSR:LFLIN? ... 5

Strany 3 - SAFETY SUMMARY

Boonton 4540 Series RF Power Meter Operation 4-52 4.5.31 Display > Graph Header Menu The Graph header menu is used to customize the text fields a

Strany 4

Boonton 4540 Series RF Power Meter Operation 4-53 Table 4-1 Field Parameter Selection List Field Parameter Description Pulse Width Automatic Pulse

Strany 5

Boonton 4540 Series RF Power Meter Operation 4-54 4.5.32 Display > Text Mode Menu The Text Mode menu can be used to configure the Modulated text

Strany 6

Boonton 4540 Series RF Power Meter Operation 4-55 4.5.33 Display > Text Mode > Ch1|2 Options Menu The Ch1|2 Options menu provides methods for

Strany 7

Boonton 4540 Series RF Power Meter Operation 4-56 Main Pg 2 > Display > Text Mode > Ch1|2 Options > Field 2 Chan Description: Select chan

Strany 8

Boonton 4540 Series RF Power Meter Operation 4-57 4.5.34 Display > Backlight Menu The Backlight menu provides methods for adjusting backlight and

Strany 9

Boonton 4540 Series RF Power Meter Operation 4-58 4.5.35 System Menu The System menu provides access to system configuration, information, and contro

Strany 10 - Contents

Boonton 4540 Series RF Power Meter Operation 4-59 4.5.36 System > I/O Config Menu The I/O Config menu provides access to set GPIB address and con

Strany 11

Boonton 4540 Series RF Power Meter Operation 4-60 4.5.37 System > I/O Config > Ethernet Menu The Ethernet menu provides access to configuratio

Strany 12

Boonton 4540 Series RF Power Meter Operation 4-61 4.5.38 System > I/O Config > Ethernet > Extensions Menu The Ethernet Extensions menu cont

Strany 13

Boonton 4540 Series RF Power Meter Contents xi READ:MARKer:CURsor:POWer? ... 5-50 READ

Strany 14

Boonton 4540 Series RF Power Meter Operation 4-62 Main Pg 2 > System > I/O Config > Ethernet > Extensions > MAC Address Description:

Strany 15 - 1. General Information

Boonton 4540 Series RF Power Meter Operation 4-63 4.5.39 System > I/O Config > Multi IO Menu This menu selects the signal to be sent to the Mu

Strany 16 - General Information

Boonton 4540 Series RF Power Meter Operation 4-64 4.5.40 System > I/O Config > Multi IO > IO Mode Status Out Menu This menu selects a TTL c

Strany 17 - 1.3 Features

Boonton 4540 Series RF Power Meter Operation 4-65 4.5.41 System > I/O Config > Multi IO > IO Mode Rec Out Menu This menu selects an analog

Strany 18

Boonton 4540 Series RF Power Meter Operation 4-66 Main Pg 2 > System > I/O Config > Multi IO > Scaling (Rec Out) Description: Sele

Strany 19 - 1.4 Accessories

Boonton 4540 Series RF Power Meter Operation 4-67 4.5.43 System > I/O Config > Multi IO > IO Mode Trig Out Menu This menu selects trigger r

Strany 20 - 1.6 Specifications

Boonton 4540 Series RF Power Meter Operation 4-68 4.5.44 System > I/O Config > Multi IO > IO Mode Volt Out Menu This menu selects a voltage

Strany 21

Boonton 4540 Series RF Power Meter Operation 4-69 4.5.45 System > Calibrator Menu (Internal 50 MHz) The Calibrator Menu is used to select and con

Strany 22

Boonton 4540 Series RF Power Meter Operation 4-70 4.5.46 System > Calibrator Menu (External 1 GHz - optional) The Calibrator Menu is used to secl

Strany 23

Boonton 4540 Series RF Power Meter Operation 4-71 Main Pg 2 > System > Calibrator > Status Description: This menu item is not implemented a

Strany 24

Boonton 4540 Series RF Power Meter Contents xii STATus:QUEStionable:EVENt? ... 5-6

Strany 25 - 2. Installation

Boonton 4540 Series RF Power Meter Operation 4-72 4.5.47 System > Calibrator > Pulse - Preset Menu (Ext 1 GHz - opt.) The pulse menu is speci

Strany 26 - 2.3 Connections

Boonton 4540 Series RF Power Meter Operation 4-73 4.5.48 System > Calibrator > Pulse - Variable Menu (Ext 1 GHz - opt.) The pulse menu is spe

Strany 27 - 2.4 Preliminary Check

Boonton 4540 Series RF Power Meter Operation 4-74 FFF Fig. 4-62.

Strany 28

Boonton 4540 Series RF Power Meter Operation 4-75 4.5.49 System > Reports Menu System Reports are text displays that provide additional informat

Strany 29 - 3. Getting Started

Boonton 4540 Series RF Power Meter Operation 4-76 Main Pg 2 > System > Reports > Info The Hardware Info Report shows temperature, power sup

Strany 30 - Getting Started

Boonton 4540 Series RF Power Meter Operation 4-77 Main Pg 2 > System > Reports > Remote Cmds The Remote Commands Report is a list of all 454

Strany 31

Boonton 4540 Series RF Power Meter Operation 4-78 4.5.50 System > Servicing Menu This menu accesses various service related functions. Main Pg

Strany 32

Boonton 4540 Series RF Power Meter Operation 4-79 4.5.51 System > Servicing > Security Menu This menu provides a means to erase sensitive stor

Strany 33

Boonton 4540 Series RF Power Meter Operation 4-80 Main Pg 2 > System > Servicing > Security > Full Sanitize Description: Select

Strany 34 - 3.3 LCD Monitor Display

Boonton 4540 Series RF Power Meter Operation 4-81 4.5.52 System > Sensor Data Menu This menu provides information about the characteristics and c

Strany 35

Boonton 4540 Series RF Power Meter Contents xiii TRIGger:POSition ...

Strany 36

Boonton 4540 Series RF Power Meter Operation 4-82 Main Pg 2 > System > Sensor Data > Sensor 2 Info Description: See a typical CW sensor re

Strany 37

Boonton 4540 Series RF Power Meter Operation 4-83 4.5.53 Setup Menu This menu provides for system preset with default parameters, pre-defined parame

Strany 38 - Caution

Boonton 4540 Series RF Power Meter Operation 4-84 4.5.54 Setup > Preset Menu This menu provides presets for widely used communications system sig

Strany 39

Boonton 4540 Series RF Power Meter Operation 4-85 4.5.55 Setup > User Presets Menu Main Pg 2 > Setup > User Preset > Preset Name Descr

Strany 40

Boonton 4540 Series RF Power Meter Operation 4-86 4.6 Cal/Zero Key Menu Pressing thefront panel Cal/Zero key wil

Strany 41

Boonton 4540 Series RF Power Meter Remote Operation 5-1 5. Remote Operation 5.1 GPIB Configuration The 4540 GPIB interface is configured using the

Strany 42 - STEP PROCEDURE

Boonton 4540 Series RF Power Meter Remote Operation 5-2 5.4 SCPI Language The 4540 Series instruments may be remotely controlled using comm

Strany 43

Boonton 4540 Series RF Power Meter Remote Operation 5-3 5.4.5 Command Arguments and Queries Many commands require arguments. In this case, the entir

Strany 44

Boonton 4540 Series RF Power Meter Remote Operation 5-4 :CALCulate:STATe? Query the state of measurement channel 1 (default chan #) :CALC:STAT? Quer

Strany 45 - 3.6 Making Measurements

Boonton 4540 Series RF Power Meter Remote Operation 5-5 5.5 Basic Measurement Information The easiest way to obtain a reading is by use of the MEASu

Strany 46

Boonton 4540 Series RF Power Meter Contents xiv 9.3 Optional Configurations ...

Strany 47

Boonton 4540 Series RF Power Meter Remote Operation 5-6 5.6 SCPI Command Reference This section contains a list of all SCPI remote commands accepted

Strany 48

Boonton 4540 Series RF Power Meter Remote Operation 5-7 *ESR? Description: Return the current value of the Standard Event Status Register, then cl

Strany 49 - 4. Operation

Boonton 4540 Series RF Power Meter Remote Operation 5-8 *OPC? Description: This command examines the OPC (Operation Complete) status fla

Strany 50 - 4.3 Numerical Data Entry

Boonton 4540 Series RF Power Meter Remote Operation 5-9 *STB? Description: Return the current value of the Status Byte register. This register ha

Strany 51 - 4.4 Sensor Calibration

Boonton 4540 Series RF Power Meter Remote Operation 5-10 *WAI Description: Wait command. This command insures sequential, non-overlapped executio

Strany 52 - 4.5 Menu Reference

Boonton 4540 Series RF Power Meter Remote Operation 5-11 5.6.2 CALCulate Subsystem The CALCulate group of the command subsystem is used to configure

Strany 53

Boonton 4540 Series RF Power Meter Remote Operation 5-12 CALCulate:LIMit:UPPer[:POWer] Description: Set or return the upper limit power level for

Strany 54

Boonton 4540 Series RF Power Meter Remote Operation 5-13 CALCulate:MATH Description: Set or return the signal source or sources combined in an ari

Strany 55

Boonton 4540 Series RF Power Meter Remote Operation 5-14 CALCulate:REFerence:COLLect Description: For the selected channel, make the current powe

Strany 56

Boonton 4540 Series RF Power Meter Remote Operation 5-15 CALCulate:STATe Description: Set or return the measurement state of the selected channel.

Strany 57

Boonton 4540 Series RF Power Meter General Information 1-1 1. General Information This instruction manual provides you with the information you need

Strany 58

Boonton 4540 Series RF Power Meter Remote Operation 5-16 5.6.3 CALibration Subsystem The CALibration group of commands is used to control automatic z

Strany 59

Boonton 4540 Series RF Power Meter Remote Operation 5-17 CALibration[:INTernal]:AUTOcal Description: Performs a multi-point sensor gain calibratio

Strany 60

Boonton 4540 Series RF Power Meter Remote Operation 5-18 5.6.4 DISPlay Subsystem The DISPlay group of commands is used to control the selec

Strany 61

Boonton 4540 Series RF Power Meter Remote Operation 5-19 DISPlay:MODUlated:TIMEBASE Description: Set or return the Modulated Mode timebase in s

Strany 62

Boonton 4540 Series RF Power Meter Remote Operation 5-20 DISPlay:PULSe:TSPAN (Note: This command is supplied for compatibility with the Model

Strany 63

Boonton 4540 Series RF Power Meter Remote Operation 5-21 DISPlay:[TEXt:]LIN:RESolution Description: Set or return the display resolution for

Strany 64

Boonton 4540 Series RF Power Meter Remote Operation 5-22 DISPlay:TRACe:VCENTer Description: Set or return the power or voltage level of the horizo

Strany 65

Boonton 4540 Series RF Power Meter Remote Operation 5-23 5.6.5 FETCh Queries The FETCh? group of queries is used to return specific measurement data

Strany 66

Boonton 4540 Series RF Power Meter Remote Operation 5-24 FETCh:ARRay:AMEAsure:TIMe? Description: Returns an array of the current automatic timing

Strany 67

Boonton 4540 Series RF Power Meter Remote Operation 5-25 FETCh:ARRay:MARKer:POWer? Description: Returns an array of the current marker measurement

Strany 68

Boonton 4540 Series RF Power Meter General Information 1-2 1.2 Description The 4540 Series RF Power Meter is the latest generation of BOON

Strany 69

Boonton 4540 Series RF Power Meter Remote Operation 5-26 FETCh:INTERval:AVERage? Description: For the specified channel, return the average power

Strany 70

Boonton 4540 Series RF Power Meter Remote Operation 5-27 FETCh:INTERval:MINimum? Description: For the specified channel, return the minimum insta

Strany 71

Boonton 4540 Series RF Power Meter Remote Operation 5-28 FETCh:MARKer:CURsor:PERcent? Description: Return the CCDF cursor y-axis position in

Strany 72

Boonton 4540 Series RF Power Meter Remote Operation 5-29 FETCh:MARKer:MINimum? Description: For the specified channel and marker, return the minim

Strany 73

Boonton 4540 Series RF Power Meter Remote Operation 5-30 FETCh:TEMPerature:AVERage? Description: Return the averaged internal temperature of

Strany 74

Boonton 4540 Series RF Power Meter Remote Operation 5-31 5.6.6 INITiate and ABORt Commands The purpose of the INITiate group of commands is to start

Strany 75

Boonton 4540 Series RF Power Meter Remote Operation 5-32 5.6.7 MARKer Subsystem The MARKer group of commands is used to configure and locate measurem

Strany 76

Boonton 4540 Series RF Power Meter Remote Operation 5-33 MARKer:POSItion:TIMe Description: Set or return the time (x-axis-position) of the selecte

Strany 77

Boonton 4540 Series RF Power Meter Remote Operation 5-34 5.6.8 MEASure Queries The MEASure group of commands is used to acquire data using a set of h

Strany 78

Boonton 4540 Series RF Power Meter Remote Operation 5-35 5.6.9 MEMory Subsystem The MEMory group of commands is used to save and recall instrument op

Strany 79

Boonton 4540 Series RF Power Meter General Information 1-3 1.3 Features Software. A 32-bit control computer running a real-time operating system

Strany 80

Boonton 4540 Series RF Power Meter Remote Operation 5-36 MEMory:SNSR:INFO? Description: Return the sensor ID and parameter data. Syntax: MEMor

Strany 81

Boonton 4540 Series RF Power Meter Remote Operation 5-37 MEMory:SYS:LOAD Description: Recall a previously stored file image of an instrument setup

Strany 82

Boonton 4540 Series RF Power Meter Remote Operation 5-38 5.6.10 OUTPut Subsystem The OUTPut group of commands is used to control various outputs of t

Strany 83

Boonton 4540 Series RF Power Meter Remote Operation 5-39 OUTPut:EXTernal:PRESent? Description: Return true if the Model 2530 external 1 GHz calib

Strany 84

Boonton 4540 Series RF Power Meter Remote Operation 5-40 OUTPut:EXTernal:PULse:PERWID Description: Set or return the pulse period <arg1> and

Strany 85

Boonton 4540 Series RF Power Meter Remote Operation 5-41 OUTPut:INTernal:SIGNal Description: Set or return the on/off state of the internal 50 M

Strany 86

Boonton 4540 Series RF Power Meter Remote Operation 5-42 OUTPut:[MIO:]RECOroderMIN Description: Set or return the Multi IO recorder output

Strany 87

Boonton 4540 Series RF Power Meter Remote Operation 5-43 OUTPut:MIO:STATus:SETTing Description: Set/return Multi IO status out setting. Below is

Strany 88

Boonton 4540 Series RF Power Meter Remote Operation 5-44 OUTPut:MIO:TRIGout:SOURce Description: Set/return Multi IO source for trigger out mode.

Strany 89

Boonton 4540 Series RF Power Meter Remote Operation 5-45 5.6.11 READ Queries The purpose of the READ? group of queries is to initiate a measurement c

Strany 90

Boonton 4540 Series RF Power Meter General Information 1-4 Dual Independent Channels. When equipped with the optional second measurement channel, t

Strany 91

Boonton 4540 Series RF Power Meter Remote Operation 5-46 READ:ARRay:AMEAsure:TIMe? Description: Returns an array of the current automatic timing m

Strany 92

Boonton 4540 Series RF Power Meter Remote Operation 5-47 READ:ARRay:MARKer:POWer? Description: Returns an array of the current marker measurement

Strany 93

Boonton 4540 Series RF Power Meter Remote Operation 5-48 READ:INTERval:MINFilt? Description: For the specified channel, return the minimum

Strany 94

Boonton 4540 Series RF Power Meter Remote Operation 5-49 READ:MARKer:AVERage? Description: For the specified channel and marker, return the averag

Strany 95

Boonton 4540 Series RF Power Meter Remote Operation 5-50 READ:MARKer:CURsor:PERcent? Description: Return the CCDF cursor y-axis position in

Strany 96

Boonton 4540 Series RF Power Meter Remote Operation 5-51 READ:MARKer:RDELTa? Description: For the specified channel return the difference between

Strany 97

Boonton 4540 Series RF Power Meter Remote Operation 5-52 5.6.12 SENSe Subsystem The purpose of the SENSe command subsystem is to directly configure d

Strany 98

Boonton 4540 Series RF Power Meter Remote Operation 5-53 SENSe:ARRay:CALTemp? Description: Returns both channels’ sensor calibration temperatu

Strany 99

Boonton 4540 Series RF Power Meter Remote Operation 5-54 SENSe:CORRection:FREQuency Description: Set or return the RF frequency for the cu

Strany 100 - Operation

Boonton 4540 Series RF Power Meter Remote Operation 5-55 SENSe:FILTer:STATe Description: Set or return the current setting of the integration filt

Strany 101

Boonton 4540 Series RF Power Meter General Information 1-5 1.4 Accessories Optional 4540 accessories that can be ordered from Boonton Electronics.

Strany 102

Boonton 4540 Series RF Power Meter Remote Operation 5-56 SENSe:PULSe:ENDGT Description: Set or return the point on a pulse, which is used to defin

Strany 103

Boonton 4540 Series RF Power Meter Remote Operation 5-57 SENSe:PULSe:STARTGT Description: Set or return the point on a pulse, which is used to def

Strany 104

Boonton 4540 Series RF Power Meter Remote Operation 5-58 5.6.13 SENSe:MBUF Data Array Commands The MBUF commands (which are grouped under the SENSe s

Strany 105

Boonton 4540 Series RF Power Meter Remote Operation 5-59 SENSe:MBUF:SIZe Description: Set or return the size of the measurement buffer.

Strany 106

Boonton 4540 Series RF Power Meter Remote Operation 5-60 SENSe:MBUF:POSition Description: Return the current acquisition (write) position into the

Strany 107

Boonton 4540 Series RF Power Meter Remote Operation 5-61 SENSe:MBUF:MEASurement Description: Set or return the desired measurement value to be

Strany 108

Boonton 4540 Series RF Power Meter Remote Operation 5-62 5.6.14 STATus Commands The STATus command subsystem enables you to control the SCPI

Strany 109

Boonton 4540 Series RF Power Meter Remote Operation 5-63 STATus:DEVice:EVENt? Description: Returns the current contents of the Device Event regist

Strany 110

Boonton 4540 Series RF Power Meter Remote Operation 5-64 STATus:OPERation:CONDition? Description: Return the current value of the Operation Condit

Strany 111

Boonton 4540 Series RF Power Meter Remote Operation 5-65 STATus:OPERation:NTRansition Description: Set or return the value of the negative transit

Strany 112

Boonton 4540 Series RF Power Meter Contents ii INSTRUCTION MANUAL, 4540 SERIES RF POWER METER Revision date 10/04/2012 © Copyri

Strany 113

Boonton 4540 Series RF Power Meter General Information 1-6 1.6 Specifications Performance specifications for the 4540 Series are listed in Table 1-2

Strany 114

Boonton 4540 Series RF Power Meter Remote Operation 5-66 STATus:QUEStionable:CONDition? Description: Return the current value of the Questionable

Strany 115

Boonton 4540 Series RF Power Meter Remote Operation 5-67 STATus:QUEStionable:EVENt? Description: Returns the current contents of the Questionable

Strany 116

Boonton 4540 Series RF Power Meter Remote Operation 5-68 STATus:QUEStionable:CALibration:CONDition? Description: Return the current value of the Q

Strany 117

Boonton 4540 Series RF Power Meter Remote Operation 5-69 STATus:QUEStionable:CALibration:NTRansition Description: Set or return the value of the neg

Strany 118

Boonton 4540 Series RF Power Meter Remote Operation 5-70 5.6.15 SYSTem Subsystem The SYSTem group of commands is used to control system-level functio

Strany 119

Boonton 4540 Series RF Power Meter Remote Operation 5-71 SYSTem:COMMunicate:LAN:ADDRess Description: Set or return the IP address for the Ethernet

Strany 120

Boonton 4540 Series RF Power Meter Remote Operation 5-72 SYSTem:COMMunicate:LAN:MAC? Description: Return the instrument’s Ethernet MAC address.

Strany 121

Boonton 4540 Series RF Power Meter Remote Operation 5-73 SYSTem:DATE:MONTH Description: Set or return the month of the real time clock. Syntax:

Strany 122

Boonton 4540 Series RF Power Meter Remote Operation 5-74 SYSTem:ERRor:CODE? Description: Returns the next queued error code number. Note t

Strany 123

Boonton 4540 Series RF Power Meter Remote Operation 5-75 SYSTem:TIMe:HOUR Description: Set or return the hour of the real time clock. Syntax:

Strany 124

Boonton 4540 Series RF Power Meter General Information 1-7 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subje

Strany 125

Boonton 4540 Series RF Power Meter Remote Operation 5-76 5.6.16 TRACe Data Array Commands The TRACe group of commands is used to control the output o

Strany 126

Boonton 4540 Series RF Power Meter Remote Operation 5-77 TRACe:COUNt Description: Set or return the number of trace points, which will be returned

Strany 127

Boonton 4540 Series RF Power Meter Remote Operation 5-78 5.6.17 TRIGger Subsystem The TRIGger group of commands is used to control synchronization of

Strany 128

Boonton 4540 Series RF Power Meter Remote Operation 5-79 TRIGger:DELay Description: Set or return the trigger delay time in seconds with respect t

Strany 129

Boonton 4540 Series RF Power Meter Remote Operation 5-80 TRIGger:MODe Description: Set or return the trigger mode for synchronizing data acquisit

Strany 130

Boonton 4540 Series RF Power Meter Remote Operation 5-81 TRIGger:SOURce Description: Set or return the trigger source used for synchroniz

Strany 131

Boonton 4540 Series RF Power Meter Remote Operation 5-82 5.6.18 SCPI Command Summary Table 5-1. SCPI COMMAND SUMMARY *CLS Clear Status Command *

Strany 132

Boonton 4540 Series RF Power Meter Remote Operation 5-83 Table 5-1. SCPI COMMAND SUMMARY (continued) CALCulate[1|2]:PKHLD Set/return state of the p

Strany 133

Boonton 4540 Series RF Power Meter Remote Operation 5-84 Table 5-1. SCPI COMMAND SUMMARY (continued) DISPlay:SCREensaver:STATe Set/ret

Strany 134 - 4.6 Cal/Zero Key Menu

Boonton 4540 Series RF Power Meter Remote Operation 5-85 5-1. SCPI COMMAND SUMMARY (continued) FETCh[1|2]:INTERval:PKAVG? Return peak to avera

Strany 135 - 5. Remote Operation

Boonton 4540 Series RF Power Meter General Information 1-8 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subje

Strany 136 - 5.4 SCPI Language

Boonton 4540 Series RF Power Meter Remote Operation 5-86 Table 5-1. SCPI COMMAND SUMMARY (continued) MEMory[1|2]:SNSR:CFFAST? Return the sensor hig

Strany 137

Boonton 4540 Series RF Power Meter Remote Operation 5-87 Table 5-1. SCPI COMMAND SUMMARY (continued) OUTPut:EXTernal:SIGNal Set or return the on/of

Strany 138

Boonton 4540 Series RF Power Meter Remote Operation 5-88 Table 5-1. SCPI COMMAND SUMMARY (continued) READ[1|2]:ARRay:AMEAsure:POWer? Return new Pp

Strany 139

Boonton 4540 Series RF Power Meter Remote Operation 5-89 Table 5-1. SCPI COMMAND SUMMARY (continued) SENSe[1|2]:CORRection:CALFactor Set co

Strany 140 - 5.6 SCPI Command Reference

Boonton 4540 Series RF Power Meter Remote Operation 5-90 Table 5-1. SCPI COMMAND SUMMARY (continued) SENSe:MBUF:COUNt Set/return number of buffer p

Strany 141

Boonton 4540 Series RF Power Meter Remote Operation 5-91 Table 5-1. SCPI COMMAND SUMMARY (continued) SYSTem:AUTOSET Perfor

Strany 142

Boonton 4540 Series RF Power Meter Remote Operation 5-92 Table 5-1. SCPI COMMAND SUMMARY (continued) SYSTem:TIMe:MINUTE Set/ge

Strany 143

Boonton 4540 Series RF Power Meter Remote Operation 5-93 This page intentionally left blank.

Strany 145

Boonton 4540 Series RF Power Meter Application Notes 6-1 6. Application Notes This section provides supplementary material to enhance your knowledge

Strany 146

Boonton 4540 Series RF Power Meter General Information 1-9 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subje

Strany 147

Boonton 4540 Series RF Power Meter Application Notes 6-2 Figure 6-1. Pulsed RF Signal Figure 6-2. Distorted Pulse Signal

Strany 148

Boonton 4540 Series RF Power Meter Application Notes 6-3 6.1.2 Diode Detection Wideband diode detectors are the dominant power sensing device used

Strany 149

Boonton 4540 Series RF Power Meter Application Notes 6-4 Frequency Response. The carrier frequency response of a diode detector is determin

Strany 150

Boonton 4540 Series RF Power Meter Application Notes 6-5 6.2 Pulse Definitions IEEE Std 194™-1977 Standard Pulse Terms and Definitions ―prov

Strany 151

Boonton 4540 Series RF Power Meter Application Notes 6-6 Table 6-1. Pulse Terms TERM DEFINITION Base Line The two portions of a pulse w

Strany 152

Boonton 4540 Series RF Power Meter Application Notes 6-7 6.3.2 Automatic Measurement Terms The following terms appear in the 4540 Series Tex

Strany 153

Boonton 4540 Series RF Power Meter Application Notes 6-8 6.3.3 Automatic Measurement Sequence The automatic measurement process analyzes the captur

Strany 154

Boonton 4540 Series RF Power Meter Application Notes 6-9 7. The process locates the bottom amplitude (baseline) using the IEEE histogram method. A h

Strany 155

Boonton 4540 Series RF Power Meter Application Notes 6-10 Item dBm mW Mesial value 10.0 10.0 Sample n 8.0 6

Strany 156

Boonton 4540 Series RF Power Meter Application Notes 6-11 12. The processor calculates the output values according to the following definitions: a

Strany 157

Boonton 4540 Series RF Power Meter General Information 1-10 Table 1-2 4540 Series Performance Specifications (continued) (Specifications are subj

Strany 158

Boonton 4540 Series RF Power Meter Application Notes 6-12 Figure 6-7 Sampling Intervals 14. The processor calculates the delay between

Strany 159

Boonton 4540 Series RF Power Meter Application Notes 6-13 Fig. 6-9 Statistical Mode Text Display Page 1

Strany 160

Boonton 4540 Series RF Power Meter Application Notes 6-14 Table 6-3. Statistical Automatic Measurements In the Statistical mode the following five a

Strany 161

Boonton 4540 Series RF Power Meter Application Notes 6-15 The following two global status values are displayed: TEXT

Strany 162

Boonton 4540 Series RF Power Meter Application Notes 6-16 6.5 Measurement Accuracy The 4540 Series includes a precision, internal, 50 MHz RF referen

Strany 163

Boonton 4540 Series RF Power Meter Application Notes 6-17 The worst case approach is a very conservative method in which the extreme conditions of eac

Strany 164

Boonton 4540 Series RF Power Meter Application Notes 6-18

Strany 165

Boonton 4540 Series RF Power Meter Application Notes 6-19 The value to use for calibration level uncertainty depends upon the sensor calibration techn

Strany 166

Boonton 4540 Series RF Power Meter Application Notes 6-20 Sensor Temperature Coefficient. This term is the error which occurs when the sensor’s tempe

Strany 167

Boonton 4540 Series RF Power Meter Application Notes 6-21 The sensor calfactor uncertainty is due to uncertainties encountered while performing this f

Strany 168

Boonton 4540 Series RF Power Meter Installation 2-1 2. Installation This section contains unpacking and repacking instructions, power requirements,

Strany 169

Boonton 4540 Series RF Power Meter Application Notes 6-22 Step 4: The Source Mismatch Uncertainty is calculated using the formula in the

Strany 170

Boonton 4540 Series RF Power Meter Application Notes 6-23 Step 10: Now that each of the individual uncertainty terms has been determined, we can combi

Strany 171

Boonton 4540 Series RF Power Meter Application Notes 6-24 Step 2: The Calibrator Level Uncertainty for the Model 2530 External 1GHz calibr

Strany 172

Boonton 4540 Series RF Power Meter Application Notes 6-25 Step 9: The Sensor Calfactor Uncertainty needs to be interpolated from the uncertainty valu

Strany 174

Boonton 4540 Series RF Power Meter Appendix A 7-1 7. Maintenance This section presents procedures for maintaining the 4540 Series. 7.1 Safety

Strany 175

Boonton 4540 Series RF Power Meter Appendix A 7-2 7.5 Firmware Upgrade Operating firmware has been loaded into the 4540 Series instrument at the fac

Strany 176

Boonton 4540 Series RF Power Meter Appendix A 7-3 How to Start Up in Boot Loader Mode 1. From a powered-down state, press and hold the to

Strany 177

Boonton 4540 Series RF Power Meter Appendix A 7-4 This page intentionally left blank.

Strany 178

Boonton 4540 Series RF Power Meter Appendix A 8-5 8. Appendix A SCPI Error Messages 8.1 SCPI Error Messages NO. MESSAGE DESCRIPTIO

Strany 179

Boonton 4540 Series RF Power Meter Installation 2-2 Table 2-1 4540 Series Packing List INSTRUMENT (See also Table 1-1) 4540 Series RF Power Meter

Strany 180

Boonton 4540 Series RF Power Meter Appendix A 8-6 NO. MESSAGE DESCRIPTION -245 "Sensor Disconnected." -246 ―Sensor voltage e

Strany 181

Boonton 4540 Series RF Power Meter Appendix A 8-7 NO. MESSAGE DESCRIPTION -342 "Temp comp pwr non-mono" -343 "Temp comp

Strany 182

Boonton 4540 Series RF Power Meter Appendix A 8-8 This page intentionally left blank.

Strany 183

Boonton 4540 Series RF Power Meter Appendix B 9-1 9. Appendix B Model 2530 1GHz Calibrator 9.1 Description The Model 2530 operates as an accesso

Strany 184

Boonton 4540 Series RF Power Meter Appendix B 9-2 9.4 Specifications RF Characteristics Operating Modes: CW, Internal or external pulse Output Fre

Strany 185

Boonton 4540 Series RF Power Meter Appendix B 9-3 9.5 Connections 1. Plug the supplied RJ-11 cable into the rear panel jack of the 2530 and into th

Strany 186

Boonton 4540 Series RF Power Meter Appendix B 9-4 This page intentionally left blank.

Strany 187

Boonton 4540 Series RF Power Meter Appendix C 10-1 10. Appendix C Boonton EULA "Boonton Electronics 4540 Series Operating Software" END

Strany 188

Boonton 4540 Series RF Power Meter Appendix C 10-2 8. U.S. GOVERNMENT LICENSE RIGHTS. All Product provided to the U.S. Government pursuant

Strany 189

Boonton 4540 Series RF Power Meter Appendix D 11-1 11. Appendix D Warranty & Repair Repair Policy 4540 Series Instrument. If the Boonton 4540

Strany 190

Boonton 4540 Series RF Power Meter Installation 2-3 Trigger Most triggered applications can use the RF signal applied to the sensors for triggering.

Strany 191

Boonton 4540 Series RF Power Meter Appendix D 11-2 THE FOREGOING WARRANTIES ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUD-ING, B

Strany 192

Boonton 4540 Series RF Power Meter Installation 2-4 6. On the front panel, press the Menu key followed by the More softkey, then on the softkeys, pre

Strany 193

Boonton 4540 Series RF Power Meter Getting Started 3-1 3. Getting Started This chapter will introduce the user to the 4540 Series. The chapter wil

Strany 194

Boonton 4540 Series RF Power Meter Contents iii SAFETY SUMMARY The following general safety precautions must be observed during all phases of operat

Strany 195

Boonton 4540 Series RF Power Meter Getting Started 3-2 Figure 3-1. Standard 4540 Series RF Power Meter - Front Panel Table 3-1 Operating Control

Strany 196

Boonton 4540 Series RF Power Meter Getting Started 3-3 Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear

Strany 197

Boonton 4540 Series RF Power Meter Getting Started 3-4 Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear

Strany 198

Boonton 4540 Series RF Power Meter Getting Started 3-5 Table 3-1 Operating Controls, Indicators and Connections (continued) Reference # Front Rear

Strany 199

Boonton 4540 Series RF Power Meter Getting Started 3-6 3.3 LCD Monitor Display This subsection includes a picture (Figure 3-3) and a table (Table 3-

Strany 200

Boonton 4540 Series RF Power Meter Getting Started 3-7 Table 3-2 LCD Monitor Display Fields (continued) Ref. No. Field Name Mode Descriptio

Strany 201

Boonton 4540 Series RF Power Meter Getting Started 3-8 Table 3-2 LCD Monitor Display Fields (continued) 10 GPIB/Remote Status All The GPIB/Re

Strany 202

Boonton 4540 Series RF Power Meter Getting Started 3-9 Table 3-2 LCD Monitor Display Fields (continued) Ref. No. Field Name Mode Descriptio

Strany 203

Boonton 4540 Series RF Power Meter Getting Started 3-10 3.4 Initialize The procedures presented in this section will initialize the 4540 Series and

Strany 204

Boonton 4540 Series RF Power Meter Getting Started 3-11 Table 3-3. Initialized Parameters Parameter Applies to: Default Graph/Text Mod

Strany 205

Boonton 4540 Series RF Power Meter Contents iv SAFETY SYMBOLS This safety requirement symbol (located on the rear panel) has been adopted by the Inte

Strany 206

Boonton 4540 Series RF Power Meter Getting Started 3-12 Table 3-3. Initialized Parameters (continued) Parameter Applies to: Default Pa

Strany 207

Boonton 4540 Series RF Power Meter Getting Started 3-13 Table 3-3. Initialized Parameters (continued) Parameter Applies to: Default Pa

Strany 208

Boonton 4540 Series RF Power Meter Getting Started 3-14 3.5 Calibration Before any pulse measurements can be acquired with the 4540 Series, a peak p

Strany 209

Boonton 4540 Series RF Power Meter Getting Started 3-15 STEP PROCEDURE 1. Select the Calibration menu soft key. Initiate autocalibration

Strany 210

Boonton 4540 Series RF Power Meter Getting Started 3-16 Figure 3-5. Menus for Sensor Calibration Procedure

Strany 211

Boonton 4540 Series RF Power Meter Getting Started 3-17 3.6 Making Measurements What you need to know To perform accurate measurements, the followin

Strany 212

Boonton 4540 Series RF Power Meter Getting Started 3-18 3.6.2 Pulse Mode For periodic or pulsed signals, it is often necessary to analyze the power f

Strany 213

Boonton 4540 Series RF Power Meter Getting Started 3-19 3.6.3 Statistical Mode Certain signals are completely random and provide no event that can se

Strany 214

Boonton 4540 Series RF Power Meter Getting Started 3-20 Note that this analysis does not depend upon any particular test signal, or upon synchronizati

Strany 215

Boonton 4540 Series RF Power Meter Operation 4-1 4. Operation This section presents the control menus and procedures for operating the 4540 Series i

Strany 216

Boonton 4540 Series RF Power Meter Contents v 1. General Information ...

Strany 217

Boonton 4540 Series RF Power Meter Operation 4-2 4.3 Numerical Data Entry Many 4540 Series parameters can be entered directly from the numeric keypa

Strany 218

Boonton 4540 Series RF Power Meter Operation 4-3 4.4 Sensor Calibration The Model 4540 Series features a built-in AutoCal process that automa

Strany 219

Boonton 4540 Series RF Power Meter Operation 4-4 4.5 Menu Reference 4.5.1 Main Menu - Pulse and Modulated Modes - Pg 1 (Top Level) The Main Menu i

Strany 220

Boonton 4540 Series RF Power Meter Operation 4-5 4.5.2 Main Menu - Pulse and Modulated Modes - Pg 2 (Top Level) The Main Menu is the topmost menu lev

Strany 221

Boonton 4540 Series RF Power Meter Operation 4-6 4.5.3 Main Menu - Statistical Mode - Pg 1 (Top Level) The Main Menu is the topmost menu level from

Strany 222

Boonton 4540 Series RF Power Meter Operation 4-7 4.5.4 Main Menu - Statistical Mode - Pg 2 (Top Level) The Main Menu is the topmost menu level from w

Strany 223

Boonton 4540 Series RF Power Meter Operation 4-8 4.5.5 Measure Menu The Measure menu contains items for controlling measurement acquisition and mode.

Strany 224

Boonton 4540 Series RF Power Meter Operation 4-9 Main > Measure > Clear Description: Clear display traces and all data buffers for CH1 and CH2.

Strany 225

Boonton 4540 Series RF Power Meter Operation 4-10 4.5.6 Channel Menu Select the specific menu for the desired measurement or reference channel if ava

Strany 226

Boonton 4540 Series RF Power Meter Operation 4-11 4.5.7 Channel 1 | 2 Menu The Channel menu contains items that affect a single measurement

Strany 227

Boonton 4540 Series RF Power Meter Contents vi 4. Operation ...

Strany 228

Boonton 4540 Series RF Power Meter Operation 4-12 Main > Channel > Channel 1 | 2 > Calibration Description: Select the calibration menu for

Strany 229 - 6. Application Notes

Boonton 4540 Series RF Power Meter Operation 4-13 4.5.8 Channel > Calibration Menu The calibration menu contains menu items to initiate multi-step

Strany 230 - Application Notes

Boonton 4540 Series RF Power Meter Operation 4-14 Main > Channel > Channel 1 | 2 > Calibration > Sel Calibrator Description: Selects the

Strany 231

Boonton 4540 Series RF Power Meter Operation 4-15 4.5.9 Channel > Extensions Pg 1 Menu (Modulated Mode) The extensions menu provides access

Strany 232

Boonton 4540 Series RF Power Meter Operation 4-16 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Filter Time Description: Set the c

Strany 233 - 6.2 Pulse Definitions

Boonton 4540 Series RF Power Meter Operation 4-17 4.5.10 Channel > Extensions Pg 1 > Corrections Menu (Modulated Mode) The corrections menu co

Strany 234 - 6.3 Automatic Measurements

Boonton 4540 Series RF Power Meter Operation 4-18 F

Strany 235

Boonton 4540 Series RF Power Meter Operation 4-19 4.5.11 Channel > Extensions Pg 2 Menu (Modulated Mode) This menu is page two of the channel exte

Strany 236

Boonton 4540 Series RF Power Meter Operation 4-20 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF f

Strany 237

Boonton 4540 Series RF Power Meter Operation 4-21 4.5.12 Channel > Extensions Pg 2 > Alarms Menu (Modulated Mode) The Alarms menu provides co

Strany 238

Boonton 4540 Series RF Power Meter Contents vii 4.5.39 System > I/O Config > Multi IO Menu ...

Strany 239

Boonton 4540 Series RF Power Meter Operation 4-22 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit Description:

Strany 240

Boonton 4540 Series RF Power Meter Operation 4-23 4.5.13 Channel > Extensions Pg 1 Menu (Pulse Mode) The extensions menu provides access to

Strany 241

Boonton 4540 Series RF Power Meter Operation 4-24 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse Description: Selects t

Strany 242

Boonton 4540 Series RF Power Meter Operation 4-25 4.5.14 Channel > Extensions Pg 1 > Corrections Menu (Pulse Mode) The corrections menu contai

Strany 243

Boonton 4540 Series RF Power Meter Operation 4-26 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle Descript

Strany 244 - 6.5 Measurement Accuracy

Boonton 4540 Series RF Power Meter Operation 4-27 4.5.15 Channel > Extensions Pg 1 > Define Pulse Menu (Pulse Mode) The define pulse menu conta

Strany 245

Boonton 4540 Series RF Power Meter Operation 4-28 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Define Pulse > Pulse Units Descrip

Strany 246

Boonton 4540 Series RF Power Meter Operation 4-29 4.5.16 Channel > Extensions Pg 1 > Define Pulse > Pulse Gate Menu Main Channel > Chan

Strany 247

Boonton 4540 Series RF Power Meter Operation 4-30 4.5.17 Channel > Extensions Pg 2 Menu (Pulse Mode) This menu is page two of the channel extensi

Strany 248

Boonton 4540 Series RF Power Meter Operation 4-31 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF fr

Strany 249

Boonton 4540 Series RF Power Meter Contents viii *SRE ...

Strany 250

Boonton 4540 Series RF Power Meter Operation 4-32 4.5.18 Channel > Extensions Pg 2 > Alarms Menu (Pulse Mode) The Alarms menu provides contro

Strany 251

Boonton 4540 Series RF Power Meter Operation 4-33 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Alarms > Low Limit Description:

Strany 252

Boonton 4540 Series RF Power Meter Operation 4-34 4.5.19 Channel > Extensions Pg 1 Menu (Statistical Mode) The extensions menu provides access t

Strany 253

Boonton 4540 Series RF Power Meter Operation 4-35 4.5.20 Channel > Extensions Pg 1 > Corrections Menu (Statistical Mode) The corrections menu

Strany 254

Boonton 4540 Series RF Power Meter Operation 4-36 Main > Channel > Channel 1 | 2 > Extensions Pg 1 > Corrections > Duty Cycle Descript

Strany 255 - 7. Maintenance

Boonton 4540 Series RF Power Meter Operation 4-37 4.5.21 Channel > Extensions Pg 2 Menu (Statistical Mode) This menu is page two of the channel ex

Strany 256 - 7.5 Firmware Upgrade

Boonton 4540 Series RF Power Meter Operation 4-38 Main > Channel > Channel 1 | 2 > Extensions Pg 2 > Frequency Description: Set the RF f

Strany 257

Boonton 4540 Series RF Power Meter Operation 4-39 4.5.22 Channel > Extensions Pg 2 > Alarms Menu (Statistical Mode) The Alarms menu provides

Strany 258

Boonton 4540 Series RF Power Meter Operation 4-40 Main > Channel > Channel 1 | 2 > Extensions Pg2 > Alarms > Low Limit Description:

Strany 259 - 8.1 SCPI Error Messages

Boonton 4540 Series RF Power Meter Operation 4-41 4.5.23 Channel > Ref 1 | 2 Menu The Reference Menu is used to manually or automatically set a p

Strany 260

Boonton 4540 Series RF Power Meter Contents ix 5.6.5 FETCh Queries ...

Strany 261

Boonton 4540 Series RF Power Meter Operation 4-42 Main > Channel > Ref 1 | 2 > Set Ref Description: Manually set the ratiometric mode ref

Strany 262

Boonton 4540 Series RF Power Meter Operation 4-43 4.5.24 Trigger Menu (Pulse Mode) The Trigger menu is used to configure trigger settings for pulse m

Strany 263 - 9.3 Optional Configurations

Boonton 4540 Series RF Power Meter Operation 4-44 Main > Trigger > Trig Slope Description: Set the trigger slope or polarity. When set to +, t

Strany 264 - 9.4 Specifications

Boonton 4540 Series RF Power Meter Operation 4-45 4.5.25 Time Menu (Modulated and Pulse Mode) The time menu provides for controlling the timebase set

Strany 265 - 9.5 Connections

Boonton 4540 Series RF Power Meter Operation 4-46 Main > Time > Trig Delay Description: Set the trigger delay time in seconds with respect to

Strany 266

Boonton 4540 Series RF Power Meter Operation 4-47 4.5.26 Stat Mode Menu (Statistical Mode) The statistical mode menu provides for controlling the

Strany 267 - 10. Appendix C Boonton EULA

Boonton 4540 Series RF Power Meter Operation 4-48 4.5.27 Term Options Menu (Statistical Mode) The statistical mode menu provides for controlling the

Strany 268 - Appendix C

Boonton 4540 Series RF Power Meter Operation 4-49 4.5.28 Cursors Menu (Statistical Mode) The statistical cursors menu controls the mode and the verti

Strany 269 - Limited Warranty

Boonton 4540 Series RF Power Meter Operation 4-50 4.5.29 Markers Menu (Modulated and Pulse Modes) The Markers menu is used to configure measurement m

Strany 270

Boonton 4540 Series RF Power Meter Operation 4-51 4.5.30 Display Menu The Display menu is the entry point for accessing and editing display configura