4530 SERIES
RF POWER METER
INSTRUCTION MANUAL
This manual is applicable to: Revision date: 07/08/2003to: Revision date: 07/08/2003
Manual P/N: 98404800C
ersions: 20021119 and later* CD P/N: 98404899C
Instrument serial numbers: ALL* Manual P/N: 98404800C
Operating Firmware Versions: 20021119 and later* CD P/N: 98404899C
*earlier firmware may not contain all capabilities listed herein
BOONTON ELECTRONICS Web Site: www.boonton.com
A subsidiary of Noise/Com -- A Wireless Telecom Group Company Email: [email protected]
25 Eastmans Road Telephone: 973-386-9696
Parsippany, NJ 07054-0465 Fax: 973-386-9191
%
Zobrazit stránku 0
1 2 3 4 5 6 ... 169 170

Shrnutí obsahu

Strany 1 - INSTRUCTION MANUAL

4530 SERIESRF POWER METERINSTRUCTION MANUALThis manual is applicable to: Revision date: 07/08/2003to: Revision date: 07/08/2003Manual P/N: 98404800Cer

Strany 2

viiiContents Boonton Electronics4530 Series RF Power MeterList of IllustrationsILLUSTRATION PAGEC-1 4530 Series RF Power Meter . . . . . . . . . . . .

Strany 3 - Contents

4-26Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterCALCulate:UNITsDescription: Set or return default programming/display unit

Strany 4 - Contents (Cont)

4-27Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationCALCulate:LIMit:UPPerDescription: Set or return the upper limit power leve

Strany 5

4-28Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterCALCulate:LIMit:CLEarDescription: Reset (clear to zero) the selected chann

Strany 6

4-29Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationMARKer:POSition:TIMeDescription: Set or return the time (position) of the

Strany 7

4-30Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterDISPlay:%OFSTDescription: Set or return the horizontal trace offset in sta

Strany 8

4-31Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationDISPlay:TXTDescription: Set the display to single channel text mode showin

Strany 9 - List of Tables

4-32Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterDISPlay:TRACe:PCTSPANDescription: Set or return the vertical span (sensiti

Strany 10 - List of Illustrations

4-33Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationDISPlay:TEXT:RESolutionDescription: Set or return the display resolution f

Strany 11 - SAFETY SUMMARY

4-34Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterTRIGger:HOLDoffDescription: Set or return the trigger holdoff time. Trigg

Strany 12 - 4530 Series RF Power Meter

4-35Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationTRIGger:SLOPeDescription: Set or return the trigger slope or polarity. Wh

Strany 13 - Limited Warranty

ixBoonton Electronics Contents4530 Series RF Power MeterSAFETY SUMMARYThe following general safety precautions must be observed during all phases of o

Strany 14

4-36Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter4.5.11 TRACe Data Array CommandsThe TRACe group of commands is used to co

Strany 15 - 1. GENERAL INFORMATION

4-37Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.5.12 SENSe:MBUF Data Array CommandsThe MBUF commands (which are grouped

Strany 16 - Chapter 1 Boonton Electronics

4-38Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSENSe:MBUF:POSitionDescription: Return the current acquisition position in

Strany 17 - Boonton Electronics Chapter 1

4-39Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.5.13 SENSe:SBUF Data Array CommandsThe SBUF commands (which are grouped

Strany 18

4-40Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSENSe:SBUF:COUNtDescription: Set or return the number of samples (power re

Strany 19

4-41Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationSENSe:HIST:COUNtDescription: Set or return the number of histogram element

Strany 20

4-42Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSENSe:CALTAB:INDEXDescription: Set or return the array index for the first

Strany 21 - 2. INSTALLATION

4-43Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationCALibration:{INTernal|EXTernal}:FIXedcalDescription: Performs a single poi

Strany 22 - Chapter 2 Boonton Electronics

4-44Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterMEMory:SYS:STOReDescription: Saves the current instrument setup to interna

Strany 23 - 3. OPERATION

4-45Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationOUTPut:EXTernal:MODulationDescription: Set or return the output modulation

Strany 24 - Chapter 3 Boonton Electronics

xContents Boonton Electronics4530 Series RF Power Meter Figure C-1 4530 Series RF Power Meter

Strany 25 - Boonton Electronics Chapter 3

4-46Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterOUTPut:RECOrder:MEASDescription: Set or return the recorder output’s measu

Strany 26

4-47Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationOUTPut:RECOrder:FASTDescription: Set or return the state of the recorder “

Strany 27

4-48Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter4.5.18 SYSTem SubsystemThe SYSTem group of commands is used to control sy

Strany 28

4-49Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationSYSTem:LIGHTDescription: Set or return the display backlight status. OFF

Strany 29

4-50Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSYSTem:COMMunicate:GPIB:ADDRessDescription: Set or return the instrument’s

Strany 30

4-51Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationSTATus:OPERation:EVENtDescription: Return the SCPI Operation Status Event

Strany 31

4-52Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSTATus:QUEStionable:CONDitionDescription: Return the current value (condit

Strany 32

4-53Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation*ESEDescription: Set or return the Standard Event Status Enable Register.

Strany 33

4-54Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter*OPCDescription: Clear or read the OPC (OPeration Complete) status flag.

Strany 34

4-55Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation*STBDescription: Return the current value of the STatus Byte register. Th

Strany 35

xiBoonton Electronics Contents4530 Series RF Power MeterRepair PolicyModel 4531 / 4532 Instrument. If the Boonton Model 4531/4532 RF Power Meter is n

Strany 36 - Table 3-3. Measurement Pages

4-56Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter*TRGDescription: Trigger a measurement. This command is equivalent to rec

Strany 37

4-57Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationCALibration[1 | 2]:{INTernal | EXTernal}:ZERO[?] Performs sensor zero offs

Strany 38

4-58Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterMEMory:SYS[1 | 2 | 3 | 4]:LOAD recall stored instrument setup from setup m

Strany 39

4-59Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationSENSe[1 | 2]:PULse:DISTal <n> set/return pulse distal level in perce

Strany 40

4-60Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSYSTem:COMMunicate:GPIB:ADDRess <n> set/return instrument’s own GPIB

Strany 41

4-61Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.6 REMOTE SENSOR CALIBRATION4.6.1 AutoCal (Peak or CW Sensors) - When m

Strany 42 - Table 3-4. Zero/Cal Menu

4-62Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterThis method is simple but has a disadvantage. You must anticipate the tim

Strany 43

4-63Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.8 SCPI EXAMPLE PROGRAM FRAGMENTS.4.8.1 Pulse Mode Example - Signal in

Strany 44

4-64Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meterd. Use an internal signal trigger to initiate successive measurements.CAL

Strany 45

4-65Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.8.2 Modulated Mode Example - Signal in: 100 microsecond wide periodic p

Strany 46

xiiContents Boonton Electronics4530 Series RF Power Meter

Strany 47

4-66Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meterb. Making successive measurements.CALC1:MODE cw Set for CW measurement mo

Strany 48

4-67Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.9 ERROR AND STATUS CODES.The following table shows codes that will be r

Strany 49

4-68Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter

Strany 50

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-15. MAKING MEASUREMENTS5.1 SENSOR TYPESThe 4530 Series RF Power Meter c

Strany 51

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-2Frequency and linearity correction factors for Boonton CW Power Sensors

Strany 52

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-3less operation allows continuous measurement of signals without the loss

Strany 53

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-4CW mode measures the average power of CW signals (and modulated signals

Strany 54

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-5Although the sample rate in statistical mode is high enough to gather a

Strany 55

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-65.4 SELECTING THE RIGHT MEASUREMENT MODESince the 4530 Series RF Power

Strany 56

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-75.4.4 Statistical Mode. Statistical Mode is only available when using a

Strany 57

Boonton Electronics Chapter 14530 Series RF Power Meter General Information1-11. GENERAL INFORMATION1.1 DESCRIPTIONThe 4530 Series RF Power Meter is

Strany 58

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-8Averaging (Optional setting for Pulse Mode) - Sets the trace averaging

Strany 59

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-95.6 SETTINGS FOR SOME COMMON SIGNAL TYPESThe following list discusses

Strany 60

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-10Meas Mode: PulseFrequency: 850 MHz (or whatever operating frequency is

Strany 61

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-11Meas Mode: PulseFrequency: 0.815 GHz (or whatever operating frequency i

Strany 62

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-12Meas Mode: PulseFrequency: 2.44 GHz (center frequency of Bluetooth band

Strany 63

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-135.6.6 Measuring HDTV. Both of the common HDTV formats (8-VSB and COFDM

Strany 64

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-14It should be noted that measurement uncertainty calculation is a very c

Strany 65

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-15The formula for calculating RSS measurement uncertainty from worst-case

Strany 66 - Table 3-5. Main Menu Summary

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-16Source Mismatch Uncertainty. This term is the mismatch error caused by

Strany 67

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-17Sensor Noise. The noise contribution to pulse measurements depends on

Strany 68

Chapter 1 Boonton ElectronicsGeneral Information 4530 Series RF Power Meter1-21.2 FEATURESMulti-mode capability Utilizes CW sensors, Peak Power senso

Strany 69

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-185.7.3 Sample Uncertainty Calculations. The following examples show cal

Strany 70

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-19sensor, and the temperature has not drifted at all, so we can assume a

Strany 71

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-20Typical Example #2: Model 57518 Peak Power SensorMeasurement condition

Strany 72

Boonton Electronics Chapter 54530 Series RF Power Meter Making Measurements5-21Step 6: The Sensor Temperature Drift Error depends on how far the temp

Strany 73

Chapter 5 Boonton ElectronicsMaking Measurements 4530 Series RF Power Meter5-22

Strany 74

Boonton Electronics Appendix A4530 Series RF Power Meter SensorsA-1Appendix A SensorsSENSORSThis appendix lists the characteristics of most common Boo

Strany 75 - 4. REMOTE OPERATION

Appendix A Boonton ElectronicsSensors 4530 Series RF Power MeterA-2Peak Power SensorsModel Frequency Dynamic Range Overload Sensor Risetime

Strany 76

Boonton Electronics Appendix A4530 Series RF Power Meter SensorsA-3CW Power SensorsModel Frequency Dynamic Range Overload Maximu

Strany 77

Appendix A Boonton ElectronicsSensors 4530 Series RF Power MeterA-4CW Power Sensors (Cont)Model Frequency Dynamic Range Overload

Strany 78

Boonton Electronics Appendix B4530 Series RF Power Meter 2530 1GHz CalibratorB-1Appendix B Model 2530 1GHz CalibratorMODEL 2530 1 GHz CALIBRATORThe Mo

Strany 79

Boonton Electronics Chapter 14530 Series RF Power Meter General Information1-31.4.2 Calibration SourcesInternal CalibratorOutput Frequency: 50 MHz ±

Strany 80

Appendix B Boonton Electronics2530 1GHz Calibrator 4530 Series RF Power MeterB-2MiscellaneousPower requirements: 90 to 260 VAC, 47 to 63 Hz, < 50 V

Strany 81

Chapter 1 Boonton ElectronicsGeneral Information 4530 Series RF Power Meter1-41.4.5 Measurement CharacteristicsMeasurements: Average Power*Maximum A

Strany 82

Boonton Electronics Chapter 14530 Series RF Power Meter General Information1-51.4.7 InterfaceVideo Output: Compressed representation of detected RF e

Strany 83

& 1998-2002, 2003 Boonton Electronics. All rights reserved.% is a registered trademark of Boonton Electronics, a subidiaryof Noise/Com, a Wirel

Strany 84

Chapter 1 Boonton ElectronicsGeneral Information 4530 Series RF Power Meter1-61.4.9 Physical SpecificationsDimensions: 3.5 inches (8.9 cm) high,8.4 i

Strany 85

Boonton Electronics Chapter 24530 Series RF Power Meter Installation2-12. INSTALLATION2.1 UNPACKING & REPACKINGThe 4530 Series RF Power Meter is

Strany 86

Chapter 2 Boonton ElectronicsInstallation 4530 Series RF Power Meter2-26. The sensors supplied with the instrument may vary widely in model number and

Strany 87

3-1Boonton Electronics Chapter 34530 Series RF Power Meter Operation3. OPERATION3.1 OPERATING CONTROLS, INDICATORS AND CONNECTIONSControls, indicato

Strany 88

3-2Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.2 KEY FUNCTION SUMMARYTable 3-1 references each operating key or connector to

Strany 89

3-3Boonton Electronics Chapter 34530 Series RF Power Meter OperationESC/Stop Key 7 Aborts any operation in progress when in Menu Mode or Zero/Cal Mode

Strany 90

3-4Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterEXT CAL CONTROL 14 An RJ-11 type modular telephone jack is used to connect the ins

Strany 91

3-5Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.3 DISPLAY FUNCTIONSThe screen display of the 4530 is divided into three sectio

Strany 92

3-6Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.4 OPERATING MODE SUMMARYThe 4530 can operate in several modes. It is possible

Strany 93

3-7Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.4.3 Graph Mode. The Graph Mode can present an oscilloscope style trace of power

Strany 94

iBoonton Electronics Contents4530 Series RF Power MeterContentsCHAPTER/SECTION PAGEList of Tables . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 95

3-8Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterFigure 3-8. Zero/Cal Mode (Example)3.5 MENU MODE OPERATION3.5.1Entry. When the

Strany 96

3-9Boonton Electronics Chapter 34530 Series RF Power Meter OperationIf a precise value is required, a special digit editing mode (Figure 3-10) may be

Strany 97

3-10Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Metera. Header. The header displays a title line and a line of text describing the cu

Strany 98

3-11Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.6.1 Entry. When the Text key is pressed, the 4530 enters Text Mode.3.6.2 Meas

Strany 99

3-12Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.7.1 Entry. When the Graph key is pressed, the 4530 enters Graph Mode.3.7.2 Mea

Strany 100 - Chapter 4 Boonton Electronics

3-13Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.8.4 Parameter Editing. The ∧ and ∨ keys are used to increment or decrement the

Strany 101 - Boonton Electronics Chapter 4

3-14Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.9 DISPLAY FORMATS3.9.1Channel Selection and Paging. Pressing the ∧ or ∨ keys

Strany 102

3-15Boonton Electronics Chapter 34530 Series RF Power Meter OperationChannel 1 Both Channels Channel 2CW Sensor Peak Sensor (Pulse Mode) Figure 3-15

Strany 103

3-16Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterChannel 1 Both Channels Channel 2CW Sensor Peak Sensor (Pulse Mode)Figure 3-17. T

Strany 104

3-17Boonton Electronics Chapter 34530 Series RF Power Meter OperationSince each sensor is different, the power meter must know the precise relationshi

Strany 105

iiContents Boonton Electronics4530 Series RF Power Meter3.3 Display Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 106

3-18Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.10.2 Zero Offset Adjustment. After a CW sensor is connected, and anytime a low

Strany 107

3-19Boonton Electronics Chapter 34530 Series RF Power Meter OperationBOONTON ELECTRONICS BOONTON ELECTRONICSMODEL 4530 MODEL 2530RF POWER METER 1GHZ R

Strany 108

3-20Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterTable 3-4. Zero/Cal MenuZero Chan (ALL SENSOR TYPES) Zero the sensor to remove t

Strany 109

3-21Boonton Electronics Chapter 34530 Series RF Power Meter OperationTable 3-4. Zero/Cal Menu - (Cont)CalSource Selects the calibrator to be used for

Strany 110

3-22Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.11.1 Measure Menu.The Measure menu contains items that control the taking of m

Strany 111

3-23Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.11.2 Channel Menu.The Channel menu contains items that affect a single measure

Strany 112

3-24Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Channel 1 | 2 > Params > FilterSet the integration time of t

Strany 113

3-25Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Channel 1 | 2 > Params > DutyCycleSets the duty-cycle in per

Strany 114

3-26Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Channel 1 | 2 > Params > Def Pulse > [ StartGate, EndGate

Strany 115

3-27Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Channel 1 | 2 > Params > ImpedanceCharacteristic impedance i

Strany 116

iiiBoonton Electronics Contents4530 Series RF Power Meter3.9 Display Formats . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 117

3-28Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Channel 1 | 2 > Display > ResolutnSelect the display resolut

Strany 118

3-29Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Channel 1 | 2 > FrDepOfst > Ofst SrcSelect which, if either,

Strany 119

3-30Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Channel 1 | 2 > Snsr Data > GainConstDisplays a sensor infor

Strany 120

3-31Boonton Electronics Chapter 34530 Series RF Power Meter OperationHorizontal Horizontal markers appear as horizontal bars, and measure the percent

Strany 121

3-32Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Trig/Time > Trig PosSelects the position of the trigger event o

Strany 122

3-33Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Trig/Time > HoldOffSets the trigger holdoff time. Trigger hold

Strany 123

3-34Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.11.5 Statisticl Menu.The Statisticl menu is used to configure data acquisition

Strany 124

3-35Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Statisticl > TrmCountSets the terminal count (sample population

Strany 125

3-36Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Calibratr > [ Int Signal, Ext Signal ]Control the on/off state

Strany 126

3-37Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.11.7 Save/Recl Menu.The Calibratr Menu is used to control both the internal, 5

Strany 127

ivContents Boonton Electronics4530 Series RF Power MeterContents (Cont)CHAPTER/SECTION PAGE4.4 Basic Measurement Operation . . . . . . . . . . . . .

Strany 128

3-38Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Utilities > Display > BacklightSelects the operating mode of

Strany 129

3-39Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Utilities > IEEE-488 > SRQ MaskSet the GPIB Service Request

Strany 130

3-40Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Utilities > Serial > ParityBitSelect whether the serial port

Strany 131

3-41Boonton Electronics Chapter 34530 Series RF Power Meter OperationManual The recorder output will generate a manually ranged signal that follows th

Strany 132

3-42Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterMain Menu > Utilities > Recorder > Calibrate > Slope AdjAdjusts the s

Strany 133

3-43Boonton Electronics Chapter 34530 Series RF Power Meter OperationMain Menu > Utilities > Sys-Tests > RecorderPerform a diagnostic test of

Strany 134

3-44Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter3.11.11 Menu Summary.The following table is a compressed summary the entire menu

Strany 135

3-45Boonton Electronics Chapter 34530 Series RF Power Meter OperationFrDepOfst . . . . . . . . . . . . . . . . . . . . Freq Dependent Offset SubmenuOf

Strany 136

3-46Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterUtilities . . . . . . . . . . . . . . . . . . . . . Instrument Utilities MenuIns

Strany 137

3-47Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.12 ERROR MESSAGES AND STATUS CODESThe following tables show the various error

Strany 138

vBoonton Electronics Contents4530 Series RF Power Meter5. MAKING MEASUREMENTS5.1 Sensor Types. . . . . . . . . . . . . . . . . . . . . . . . . . . .

Strany 139

3-48Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power MeterTable 3-8. Sensor Zero / Cal Status Codes00 operation complete, no errors03 coar

Strany 140

3-49Boonton Electronics Chapter 34530 Series RF Power Meter Operation3.13 RECORDER OUTPUT CALIBRATIONThe recorder output may be user calibrated for

Strany 141

3-50Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter6. The slope adjustment has no effect on the output voltage when the force settin

Strany 142

3-51Boonton Electronics Chapter 34530 Series RF Power Meter Operationmain software version, and a DSP software version. They may not be the same, but

Strany 143 - 5. MAKING MEASUREMENTS

3-52Chapter 3 Boonton ElectronicsOperation 4530 Series RF Power Meter

Strany 144 - Chapter 5 Boonton Electronics

4-1Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4. REMOTE OPERATION4.1 GPIB CONFIGURATION.The 4530 Series GPIB interface i

Strany 145 - Boonton Electronics Chapter 5

4-2Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Metercode. Upon receiving this code, the instrument immediately transmits the s

Strany 146

4-3Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationCommand which takes numeric argument: SENSe1:AVERage <n>Command with

Strany 147

4-4Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter4.5 COMMAND REFERENCE.This section contains a list of all remote commands

Strany 148

4-5Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationMEASure:VOLTageDescription: Return average voltage using a default instrume

Strany 149

viContents Boonton Electronics4530 Series RF Power MeterContents (Cont)CHAPTER/SECTION PAGEAPPENDIX AAvailable Sensors . . . . . . . . . . . . . . .

Strany 150

4-6Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterABORtDescription: Terminates any measurement in progress and resets the sta

Strany 151

4-7Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationFETCh:ARRay:MARKer:POWerDescription: Returns the current power readings at

Strany 152

4-8Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterFETCh:ARRay:PULse:POWerDescription: Returns an array of the current marker

Strany 153

4-9Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationFETCh:ARRay:AMEAsure:POWerDescription: Returns an array of the current auto

Strany 154

4-10Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterREAD:MARKer:POWerDescription: Performs a single measurement and returns th

Strany 155

4-11Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationREAD:ARRay:CW:VOLTageDescription: Performs a single measurement and return

Strany 156

4-12Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterREAD:ARRay:AMEAsure:POWerDescription: Performs a single measurement and re

Strany 157

4-13Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationCH1Description: Configures native mode talk instructions to return Channel

Strany 158

4-14Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterTKSDATADescription: Returns the sensor data table from sensor’s EEPROM. N

Strany 159

4-15Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationTKSCWRGDescription: Returns the CW sensor linearity calibration table from

Strany 160

viiBoonton Electronics Contents4530 Series RF Power MeterList of TablesTABLE PAGE3-1 Keyboard Controls, Indicators and Connectors . . . . . . . . . .

Strany 161

4-16Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterTKPWRDescription: Sets the talkmode to return the average, maximum and min

Strany 162

4-17Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationTKMK2Description: Sets the talkmode to return the reading at marker 2 for

Strany 163

4-18Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterTKPLSPWRDescription: Sets the talkmode to return an array of automatic pow

Strany 164

4-19Boonton Electronics Chapter 44530 Series RF Power Meter Remote Operation4.5.6 SENSE SubsystemThe purpose of the SENSe command subsystem is to dir

Strany 165 - Appendix A Sensors

4-20Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSENSe:FILTer:TIMeDescription: Set or return the current length of the inte

Strany 166 - Peak Power Sensors

4-21Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationSENSe:CORRection:FDOFfsetDescription: Set or return the name of the curren

Strany 167 - CW Power Sensors

4-22Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power MeterSENSe:PULSe:UNITsDescription: Set the units for entering the pulse distal,

Strany 168 - CW Power Sensors (Cont)

4-23Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationSENSe:PULSe:STARTGTDescription: Set or return the point on a pulse, which

Strany 169 - Specifications

4-24Chapter 4 Boonton ElectronicsRemote Operation 4530 Series RF Power Meter4.5.7 CALCulate SubsystemThe CALCulate group of the command subsystem is

Strany 170 - RJ-11 Telephone Cable

4-25Boonton Electronics Chapter 44530 Series RF Power Meter Remote OperationCALCulate:DCYCDescription: Set or return the pulse duty cycle in percent o