7.4.3 Sweep Mode
IM products in linear environment are frequency independent. Many passive
components show a frequency response that is less linear or strongly frequency
dependent. The sweep mode increments / decrements carrier signals by 1 MHz.
covering a Tx range that results in a sweep of the complete Rx frequency range.
Any deviations of PIM measurements that are frequency dependent are
immediately visible.
Element
Name
Description / Display
A
Pull-Down menu
Allows Screenshots, data logging, and self test. For
more information refer to chapter Pull-Down
menus”.
B
Carrier display
Carrier signal information: Frequencies, Power
levels
C
Graphical display
Shows two traces (red / yellow) of an IM product
that are generated by
a) sweeping the lowest frequency upwards red
trace and after that
b) sweeping highest frequency carrier signal